Checkline 3000FX Series User Manual
Page 12
- 12 -
4 .0 C
ALI BRAT I ON
4.1 Measuring Using Factory (Works) Calibration
1. Press ▼. Calibration appears on the display.
2. Press
OK
. Zero Setting appears on the display.
3. Press ▼ repeatedly until Delete Calibration appears.
4. Press
OK
. Works Calibration Activated appears
briefly, then the start screen with - - - -. You can now
carry out the measurement.
4.2 One-point calibration (zero procedure)
This procedure requires an uncoated test object having
similar dimensions and material properties to those of the
coated object to be measured.
1. Press ▼. Calibration appears on the display.
2. Press
OK
. Zero Setting appears.
3. Place the Measuring Sensor (item G, page 4) on the test
object at a right angle. Press gently and hold until chime
sounds, then lift up. Repeat several times.
4. Press
OK.
Zero has been set appears briefly on the display,
followed by the start screen with the four dashes - - - -.
You can now carry out the measurement.
4.3
Two-point calibration (using calibration foil)
As with Zero setting, this procedure requires an uncoated test
object having similar dimensions as the coated object that will
be measured. Choose a calibration foil that is closest to the
expected coating layer thickness.
- 17 -
3. Press ▼ until Display of Single Values appears.
To continue measuring, press
ESC
repeatedly until the start
screen with the four dashes - - - - is displayed.
To print single values:
1. Press ▼ repeatedly until Statistics appears on display.
2. Press
OK
. Printout Of Statistics appears
3. Press ▼ repeatedly until Printout of Single Values
appears
4. Point instrument with the IR transmitter (top end edge)
towards the IR adapter installed on the printer. The distance
should be approx. 12” to 20” (
30cm
to
50cm
)
5. Press
OK
. The data are transmitted, and Printing... will
briefly appear, followed by the start screen with the four
dashes - - - -.
To continue measuring, press
ESC
repeatedly until the start
screen with the four dashes - - - - is displayed again.
6.3 Display and print statistical values
To display statistical values:
1. Press ▼ repeatedly until Statistics appears on display.
2. Press
OK.
Printout Of Statistics appears.
3. Press ▼ repeatedly until Display Of Statistics appears.
4. Press
OK
repeatedly. Each time OK is pressed, a different
statistical value will be displayed in this order: N, x, s,
and then N, Max, Min. At the conclusion, Display of
Statistics appears again and the sequence of statistics
displayed will repeat.
- 12 -
4 .0 C
ALI BRAT I ON
4.1 Measuring Using Factory (Works) Calibration
1. Press ▼. Calibration appears on the display.
2. Press
OK
. Zero Setting appears on the display.
3. Press ▼ repeatedly until Delete Calibration appears.
4. Press
OK
. Works Calibration Activated appears
briefly, then the start screen with - - - -. You can now
carry out the measurement.
4.2 One-point calibration (zero procedure)
This procedure requires an uncoated test object having
similar dimensions and material properties to those of the
coated object to be measured.
1. Press ▼. Calibration appears on the display.
2. Press
OK
. Zero Setting appears.
3. Place the Measuring Sensor (item G, page 4) on the test
object at a right angle. Press gently and hold until chime
sounds, then lift up. Repeat several times.
4. Press
OK.
Zero has been set appears briefly on the display,
followed by the start screen with the four dashes - - - -.
You can now carry out the measurement.
4.3
Two-point calibration (using calibration foil)
As with Zero setting, this procedure requires an uncoated test
object having similar dimensions as the coated object that will
be measured. Choose a calibration foil that is closest to the
expected coating layer thickness.
- 17 -
3. Press ▼ until Display of Single Values appears.
To continue measuring, press
ESC
repeatedly until the start
screen with the four dashes - - - - is displayed.
To print single values:
1. Press ▼ repeatedly until Statistics appears on display.
2. Press
OK
. Printout Of Statistics appears
3. Press ▼ repeatedly until Printout of Single Values
appears
4. Point instrument with the IR transmitter (top end edge)
towards the IR adapter installed on the printer. The distance
should be approx. 12” to 20” (
30cm
to
50cm
)
5. Press
OK
. The data are transmitted, and Printing... will
briefly appear, followed by the start screen with the four
dashes - - - -.
To continue measuring, press
ESC
repeatedly until the start
screen with the four dashes - - - - is displayed again.
6.3 Display and print statistical values
To display statistical values:
1. Press ▼ repeatedly until Statistics appears on display.
2. Press
OK.
Printout Of Statistics appears.
3. Press ▼ repeatedly until Display Of Statistics appears.
4. Press
OK
repeatedly. Each time OK is pressed, a different
statistical value will be displayed in this order: N, x, s,
and then N, Max, Min. At the conclusion, Display of
Statistics appears again and the sequence of statistics
displayed will repeat.