Rs-232, Rs232 transfer protocol, Rs232 datastream format – Checkline MTM User Manual
Page 6: Rs232 cable pinouts
AWS-MTM-DP Guide
OPERATION 5
RS-232
The MTMDP display can be connected to a printer, computer or data collector via its
RS232 interface. Every time a reading is accepted into memory, a peak is cleared, or
data is transmitted via the print data menu(s), it is transmitted via the RS-232 port. To
download the readings, go to the DATA MENU.
SPECIAL NOTE: Model MTMDP-2S PTEST is not equipped with memory, so the operation
of the RS-232 will vary from the standard unit. The protocol, however, remains the same.
RS232 Transfer Protocol
Protocol
Value
Cable
9 pin to mini-plug.
Baud
9600
Parity
Flow
None
None
Bits
8
S Bit
1
RS232 Datastream Format
mmmbsdddddbuuuuucl,
where:
m
Memory Location
s
Sign (space or -)
c
Carriage Return
d
Data with Decimal Point
l
Line Feed
u
Units
b
Blank
RS232 Cable Pinouts
Pin #
Description
Pin #
Description
1 Unused
6 Unused
2
Transmit
7
Unused
3 Receive
8 Unused
4
Unused
9
Unused
5 Ground