background image

AWS-MTM-DP Guide

OPERATION 5




RS-232

The MTMDP display can be connected to a printer, computer or data collector via its
RS232 interface. Every time a reading is accepted into memory, a peak is cleared, or
data is transmitted via the print data menu(s), it is transmitted via the RS-232 port. To
download the readings, go to the DATA MENU.

SPECIAL NOTE: Model MTMDP-2S PTEST is not equipped with memory, so the operation
of the RS-232 will vary from the standard unit. The protocol, however, remains the same.

RS232 Transfer Protocol

Protocol

Value

Cable

9 pin to mini-plug.

Baud

9600

Parity

Flow

None
None

Bits

8

S Bit

1

RS232 Datastream Format

mmmbsdddddbuuuuucl,

where:

m

Memory Location

s

Sign (space or -)

c

Carriage Return

d

Data with Decimal Point

l

Line Feed

u

Units

b

Blank

RS232 Cable Pinouts

Pin #

Description

Pin #

Description

1 Unused

6 Unused

2

Transmit

7

Unused

3 Receive

8 Unused

4

Unused

9

Unused

5 Ground