beautypg.com

Power-on programming – Checkline ZP User Manual

Page 7

background image

page 7

page 14

Power-On programming

Turn on the gauge. Press

and

for 3 seconds to enter

Power-On programming (CF9 flashes with solid F0). Press

or

to cycle CF9 F0, CF9 F1, CF9 F2, CF9 F3, CF9 F4, CF9 F5,

CF9 F6 and CF9 End. Press

to select a function.

CF9 F0

Memory recall

Press SEND, memory location and value cycle. Press PEAK or ZERO to increase or
decrease memory number (and corresponding value).

CF9 F1

High & low setpoints

Default=‘0’ both Hi and Lo

Press SEND, –HI– displays, then the high set value (i.e. H 10.0). Press PEAK to
increase and ZERO to decrease, press SEND to select. –LO– displays then the low set
value (i.e. L 5.0). Press PEAK to increase and ZERO to decrease, press SEND to select,
CF9 End displays. Press SEND again to exit.

CF9 F2

Peak mode

Default=‘Or’

Press SEND, ‘Or’ or ‘And’ displays. Press PEAK or ZERO to cycle. Press SEND to
select. ‘Or Peak’ records the Peak in either compression or tension during test. ‘And
Peak’ records both the compression peak and tension peak during a test.

Zero reset memory store

Default=‘AA-OFF’

Auto memory displays after Peak mode is selected. ‘AA-On’ enables automatic memory
storage and reset to zero. ‘AA-OFF’ turns off auto function. Press PEAK or ZERO to
change. Press SEND to select, CF9End displays, press SEND again to exit.

CF9 F3

Auto zero reset

Default=‘Ac-OFF’

Press SEND, ‘Ac-On’ or ‘Ac-OFF’ displays. Press PEAK or ZERO to cycle. If ‘AC-On’ is
selected, auto zero reset duration can be programmed. Press PEAK to increase or
ZERO to decrease. Press SEND to select, CF9End displays, press SEND again to exit
(i.e. ‘SEC 3.0’ is displayed for auto zero reset duration of 3 seconds).

CF9 F4

Audible beep

Default=‘Sd-On’

Press SEND, ‘Sd-On’ for alarm on, or ‘Sd-OFF’ for off displays. Press PEAK or ZERO to
cycle. Press SEND to select. CF9End displays, press SEND again to exit (alarm sounds
for values over HI or under LO setpoints).

Setpoint alarm

Default=’’

Setpoint alarm displays after audible beep is selected. ‘AL-On’ for setpoint alarm on or
‘AL-OFF’ for off. Press PEAK or ZERO to cycle. Press SEND to select, CF9End displays,
press SEND again to exit.

CF9 F5

Reverse display

Press SEND, ‘-12345’ for standard or ’

’ for reverse displays. Press PEAK or

ZERO to cycle. Press SEND to select. Flashing CF9 with solid End displays. Press
SEND again to exit. (for vertical mounting).

CF9 F6

Auto power off duration

Default=‘AO-10’

Press SEND, ‘AO-10’ displays. Press PEAK or ZERO to cycle ‘AO-5’ for 5 min auto
power off duration, ‘AO-10’ for 10 min, ‘AO-30’ for 30 min, ‘AO-60’ for 60 min and ‘AO-
OFF’ to by-pass auto power off. Press SEND to select, CF9End displays, press SEND
again to exit.

-12

34

5

Downloading Data from Memory

Choose between the following download methods.

RS-232C Data Download from Memory
Connect the gauge and device receiving data with a CB-204
cable. Use the I[CR] ASCII command to transmit data (uppercase
ASCII character format).

USB Data Download from Memory
Use ZP Logger software to downoad data from memory using
USB, see pages 16-19.

Clearing Data from Memory

1. Turn off the gauge.

2. Press

again while holding

to enter memory mode.

Single Memory Clear

A memory location number that is displayed with a dot at both

ends is the last stored data and is the only one that can be erased.

Press

to erase and ErASEd is displayed. If you attempt to erase

any other memory location Error is displayed. If all data is erased,

– – – – – is displayed then flashing CF9 with solid End. Continue

Power-On programming or press

to exit (See page 18).

All Memory Clear

While a memory value and location are displayed, press

for 3

seconds, all data is erased and ErASEd is displayed, then – – – – –.

Continue Power-On programming or press

to exit.

Recalling Data from Memory

1. Turn on the gauge.

2. Press

and

for 3 seconds to enter Power-On program-

ming (CF9 flashes with solid F0). Press

and the display

cycles memory location and value. Press

to increase loca-

tion and

to decrease. Press

to exit.