Checkline ZP User Manual
Page 10
page 10
page 11
3. External Switch Display Freeze
By connecting #10 and #12 of the
communications port, the gauge
instantaneously captures the criti-
cal reading and holds the display
from remote locations.
4. External Switch Display Clear
By connecting #8 and #12 of the communications port, dis-
play can be cleared from remote locations.
(1) Pay extra attention to avoid overload as display
value will not change during display hold.
(2) Use contact closure only and DO NOT apply
voltage across #10 and #12 port pins.
Use contact closure only and DO NOT apply voltage
across #8 and #12 port pins.
2. ±2 VDC Analog Signal
Connect the CB-104 analog cable to the communications port
and the device receiving the data.
RECHARGING Ni-MH BATTERY
1. To maximize battery life, power automatically shuts off after
10 minutes of non-use or user-defined interval. Auto shut off
is bypassed during USB output or when using the AC
adapter/charger.
2. The battery icon flashes when the gauge needs
to be recharged. It takes 10 hours to charge fully.
3. Push
to turn off power. Only use the IMADA
AC adapter/ charger provided, AD120 for
115VAC, AD230 for 230VAC. Plug into the correct AC output.
4. When the gauge is turned off, make sure the AC adapter/charger
is disconnected to avoid overcharging.
Optional Cables
CB-104 10' Analog cable
CB-204 10' RS-232C cable, 9 pin female
Optional Handle
The OH-1 fits ZP gauges. Constructed of high qual-
ity steel for rugged use, the handle facilitates
measurement of heavy loads. Complete with
mounting screws.
NG
00.0
peak
N
OK
+NG
OV
ER
LO
AD
ZER
O
SEND
PEAK
ON
OFF
Optional Adapter Plate
AP-001 Adapter Plate mounts IMADA
low capacity gauges to most other
brands of test stands.
Use the 4 screws (supplied) to mount
the IMADA gauge to the AP-001
adapter plate. Then use the 2 PEM
nuts on the AP-001 adapter plate to
mount to other brands of test stand.
OTHER BRAND TEST STAND
2 MOUNTING PLATE BOLTS
GAUGE MOUNTING
SCREWS (4)
PEM NUTS
IMADA GAUGE
AP-001
FORCE CONTROL
Connect the Imada Z Series force gauges to an MX-
Series test stand and enable force controlled testing.
Force control can automatically stop testing when
the force value reaches the high setpoint programmed
on the Z Series force gauge, or maintain the force
between the High and Low setpoints. Force control
is ideal for fatigue or non-destructive testing.
MX-SERIES
TEST STAND
WITH ZP GAUGE
Optional Data Acquisition Software
ZP Recorder captures and processes 1,000 data/second using a USB
port.A graph of the data and statistics are displayed.Analyze all or part
of the data. Compare multiple graphs.
SW-1 software captures and analyzes force data using an RS-232 port.
A running log of data, graph and statistics are displayed.