beautypg.com

Nadia, Temperature – BUCHI NIRWare User Manual

Page 310

background image

304

NIRWare Suite

NIRWare 1.5 Software Manual

Temperature

Several temperature sensors are provided within the spectrometer. Temperature is checked at the
following positions against reference values: wedge drive, instrument, laser and main board and,
depending on the measuring cell, detector, detector board, application board. For every sensor a lower
and upper limit value is predefined. The SST for temperatures is passed, when all temperatures are
between the lower and the upper value.

The

SST test

can be performed manually or configured to be started automatically.

All results of the SST are documented in the

SST report

.

6.10.2. NADIA

NADIA stands for „NIRWare Automatic Diagnose“ and is a service tool to examine the basic functions of
the BUCHI NIRFlex-N500.

It will be used from Service Technicians to evaluate malfunctions of the instrument or to do a performance
check of the instrument as part of a service contract.