NIRWare Suite

NIRWare 1.5 Software Manual

Modifying parameters

Click the "Edit" icon

and change the application name, the application description and for qualitative

applications the hitlist size.

The application parameters are grouped in 4 different categories.

Properties

Instrument

Report

Operator Configuration

Each of this categories contain basic and advanced settings which can be modified.

6.2.2. Comparing external reference spectra and defining the

tolerance limit

(De)activating the test

The test can be activated or deactivated for each individual application (default: active). This is done in the
Management Console

– Application Designer - - Instrument: External

Reference Spectra Comparison.

Changing the tolerance limit

– Who is allowed to?

The tolerance limit of the test is found under Management Console

– Application Designer -

- Instrument: Maximum allowed deviation of external reference spectra [%], and
the default value is 5%. It is recommended not to change this value, and therefore even administrators