BUCHI NIRWare User Manual
Page 162

156
NIRWare Suite
NIRWare 1.5 Software Manual
Position
Quantitative and Identification:
Position of the parameter to be displayed in NIRWare Operator and all
printing documents.
Only needed if more than one property / property group is added to the
application. predicted.
Decimal Places
(Result)
Quantitative:
Define, how many positions after decimal point will be displayed for the
prediction result.
Default: 2
Decimal Places
(Average)
Quantitative:
Define, how many positions after decimal point will be displayed for
the result average.
Default: 2
Decimal Places (Std
Dev)
Quantitative:
Define, how many positions after decimal point will be displayed for
the standard deviation.
Default: 3
Upper action limit
Entry for the upper action limit.
"Action Limits" define the limits, which if exceeded or under-run, indicate
that results are outside of acceptable tolerance.
See also "Definition of calibration range, warning and action limits".
Default: upper limit of calibration range
To set back the default value for the action limit, click on this button
.
Upper warning limit
Entry for the upper warning limit.
"Warning Limits" define the range around an acceptable result (e.g.
production optimum).
See also "Definition of calibration range, warning and action limits".
Default: upper limit of the calibration range
To set back the default value for the warning limit, click on this button
.
Lower warning limit
Entry for the lower warning limit.
"Warning Limits" define the range around an acceptable result (e.g.
production optimum).
See also "Definition of calibration range, warning and action limits".
Default: lower limit of calibration range
To set back the default value for the warning limit, click on this button
.
Lower action limit
Entry for the lower action limit.
"Action Limits" define the limits, which if exceeded or under-run, indicate
that results are outside of acceptable tolerance.