7 | Maintenance

BÜCHI Labortechnik AG

34/46

Ultrasonic Controller

4.

Set pump speed to 8 % and switch on pump.

5.

Switch on Ultrasonic Nozzle and set power to 2 W.

6.

Turn in the adjusting screw (4) until the water is sucked in steadily.

7.

Set the correct atomization power/stall point.

• 6.1 Stall point, determination [

}

page 31]

8.

Check if turning out the adjusting screw by ¼ turn helps to decrease pul­
sation, otherwise leave the position of the previous step.

9.

Switch off pump.

• The feed must stop.

10. Switch off Ultrasonic Nozzle.

Fig. 13: Adjustment of peristaltic pump

Follow­on tasks

1.

Install spraying nozzle into the Mini Sprayer B­290.