BUCHI Ultrasonic Package User Manual
Page 34

7 | Maintenance
BÜCHI Labortechnik AG
34/46
Ultrasonic Controller
4.
Set pump speed to 8 % and switch on pump.
5.
Switch on Ultrasonic Nozzle and set power to 2 W.
6.
Turn in the adjusting screw (4) until the water is sucked in steadily.
7.
Set the correct atomization power/stall point.
• 6.1 Stall point, determination [
page 31]
8.
Check if turning out the adjusting screw by ¼ turn helps to decrease pul
sation, otherwise leave the position of the previous step.
9.
Switch off pump.
• The feed must stop.
10. Switch off Ultrasonic Nozzle.
Fig. 13: Adjustment of peristaltic pump
Followon tasks
1.
Install spraying nozzle into the Mini Sprayer B290.