beautypg.com

Ii. specifications – Brookfield Temperature Controller, Model 106 User Manual

Page 4

background image

Brookfield Engineering Labs., Inc.

Page 4

Manual No. M/02-207-C009

II.

SPECIFICATIONS

Utilities

VOLTAGE SPECIFICATIONS

Input Voltage

85 to 265 VAC

Input Frequency

50/60 Hz

Replaceable Fuses

Two fuses, 2A, 250V, 5 x 20 mm, Fast Acting

Load

TRIAC (250 watts max)
(Check the label beneath the Programmable

Temperature Controller for the voltage require-

ments of our unit)

Table II.1

Controller

MEASUREMENT/CONTROL SPECIFICATIONS

Range

15°C above ambient to 300°C

27°F above ambient to 572°F

Resolution

0.1 °C or °F

Reading Accuracy

±1.0°C (between ambient and +150°C)

±2.0°C (between +151°C and +300°C)

Setpoint Accuracy

The temperature will be maintained within 0.3°C

of the setpoint

Recorder Output Voltage

0 to 4 Volts

1 volt = 0°C/32°F; 4 Volts = 300°C/572°F

Table II.2

NOTE:

The temperature accuracies stated above are a result of the combined accuracies of

the Programmable Temperature Controller, the temperature probe, and the Thermo
container.

Thermo Container

THERMO CONTAINER SPECIFICATIONS

Range

15°C above ambient to 300°C

Accuracy

±0.5% of the controller setpoint

Table II.3