Brookfield Temperature Controller, Model 106 User Manual
Page 15
Brookfield Engineering Labs., Inc.
Page 5
Manual No. M/02-207-C009
Temperature/Time programs remain in memory when the Programmable Temperature Controller
is turned off. It is not necessary to re-enter these programs when the unit is turned on again.
1) Use the °F/°C key to select whether temperatures will be entered in the Fahrenheit or
Centigrade format. Since Table VI.1 indicates temperatures in °C, we press the °F/°C
key until a C is shown in the right most display position.
2) Press the PROGRAM key. The PROGRAM LED will illuminate and the last tempera-
ture programmed in program slot zero will be displayed as follows:
40.00
where 40 is the previously set temperature for the first temperature entry and could be any
value. If the program in Table 7 was being reviewed, 40 would be 100.
0 indicates program step 1. Note that the ten (10) available programs are numbered from
0 to 9.
3) For this example, we will assume that we are entering a program to control a Thermo
Container. The program we will enter is depicted in Table 7. We now enter the desired
temperature for the first program temperature step using the ARROW keys.
4) The display now reads:
100.00
We have now altered the current program step to correspond to the first entry for Table 7.
5) Press the SET key to accept this value. The “hold” time is now displayed for this step.
Use the arrow keys to change the displayed time to 5 minutes as listed in Table VI.1.
5 t
6) Press the SET key to accept this value and move to the next program step.