Hold, Stop – Brookfield Temperature Controller, Model 106 User Manual
Page 16
Brookfield Engineering Labs., Inc.
Page 6
Manual No. M/02-207-C009
7) 45 is the previously set temperature for the second entry.
1 indicates the second program entry.
45.01
Enter the desired temperature for the second entry (140) and press the SET key when the correct
value has been entered. Enter the desired time for the second step and press the SET key to ac-
cept it and move to the next step.
8) Continue to enter additional steps for the temperature/time program. The Programmable
Temperature Controller can accept up to 10 entries. The tenth entry is indicated by the
integer 9.
NOTE:
No matter how many steps you have in a program you will always have to press the
SET key to cycle through all ten (10) temperature time steps. This also holds if you
simply want to review a program.
9)
Press the
SET
key once to accept the last (i.e. tenth step) hold time interval and to ad-
vance to the
HoLd/StoP
decision screen:
HoLd
This could just as easily have been displaying the word
StoP
if that were the last selected
HoLd/StoP
decision. It will always on first entry display the last option selected. Here, you
are being asked if you want the Programmable Temperature Controller to maintain or terminate
the last programmed temperature (290 in our example) when the last program step time interval
(2 minutes) expires. In the terminate temperature case, the last screen would have displayed:
StoP
At this point you may select either option by using the ARROW keys to display the method of
your choice. If
StoP
is displayed, pressing the DOWN ARROW key will cause
HoLd
to be
displayed. You can return to
StoP
by pressing the UP ARROW key. Make your choice and
then press the SET key to accept it.
10) You will now see the following screen asking for your input for the start step number for
your program: