Welch Allyn GCX M-Series Wall Mount Kit for Welch Allyn Propaq CS Monitor - Installation Guide User Manual
Page 5

DU-PR-0013-06 Rev A 2/28/07 GCX Corp Page 5 of 5
Adjusting Tilt and Tilt Tension
Tilt Adjustment: Turn Tilt Adjustment Lever counter clockwise to loosen. Lever provides ratchet-style adjustment (see
Installation Note below). Lift and rotate lever to desired position and release, then loosen. Grasp the top and/or bottom
of the device and tilt to the desired angle. Turn Tilt Adjustment Lever clockwise (using ratchet feature as desired) to
tighten and lock position.
Tilt Tension: Adjust the overall tilting tension by evenly tightening or loosening the two (2) Tilt Tension Adjustment
Screws with the 5/32'' hex wrench provided. Once the overall tilt tension is set, use the Tilt Adjustment Lever to fine tune
and lock the tilt position.
Periodic Maintenance and Cleaning
All fasteners associated with the mounting system should be inspected periodically and tightened as necessary.
The mounting system may be cleaned with any non-abrasive solution as approved by the facility in which the product will
be used.
Tilt Tension Adjustment Screws (2)
Tilt Adjustment Lever
Installation Note: The Tilt Adjustment Lever is
a multi-position clamping lever that operates by
lifting, rotating, and releasing the handle.