1. Connect a USB cable to the service test box; connect the opposite end to a PC.
2. Connect a mini B USB cable to the ProBP 3400 and the opposite end to the service test box.

3. Connect the blood pressure hose to the NIBP module and the test volume.

Connect a piece of tubing to the squeeze bulb with bleeding screw; connect the opposite end to the square
fitting connector.

Connect another piece of tubing to the square fitting connector; connect the opposite end to the calibrated
pressure meter.

Connect another piece of tubing to the square fitting connector; connect the opposite end to the test volume.

Connect the last piece of tubing to the square fitting connector; connect the opposite end to the "Y" fitting on
the two-tube blood pressure connector.

OpenTopic | ProBP 3400 calibration and verification setup | 9