ProBP3400 error codes

The following tables list the error messages associated with the Welch Allyn ProBP 3400.

Error code

Related System

Additional details

000000001

Persistent storage error

EERPOM is corrupt or invalid

000000001

BP module failure

Mod F NIBP board failure

Error condition

Definition / Description

Required action

Low battery 5 minutes remaining.

The battery is extremely low. Initial
alarm condition occurs when 5
minutes of battery power remain; this
is the second low battery alarm.

Click the OK button to dismiss the
message.

Plug the device into the outlet to
dismiss the message.

Low battery 30 minutes remaining.

The battery is extremely low. Initial
alarm condition occurs when 30
minutes of battery power remain; this
is the first low battery alarm.

Click the OK button to dismiss the
message.

Plug the device into the outlet to
dismiss the message.

Powering down. Call for service.

This message could be caused by
various internal errors, power
manager, and battery faults.

The device will auto power off.

Battery is absent or faulty.

Battery is absent or faulty.

Plug the device into the outlet to
dismiss the message.

User cancelled NIBP reading.

Blood pressure reading cancelled by
user.

Retake measurement.

Inflation too quick; check NIBP cuff
and tubing connections.

The blood pressure cuff inflated too
quickly.

Check for kinked hose.

Cuff pressure limits exceeded.
Powering down.

Overpressure. Exceeded maximum
limit for this patient mode.

Limit patient movement.

NIBP air leak; check cuff and tubing
connections.

Major air leak.

Check connections (device will
power down).

Unable to determine NIBP; check
connections; limit patient movement.

Excessive patient movement.

Call Welch Allyn service.

NIBP not functional. Call for service.

Auto zero failure.

Unable to determine NIBP; check
inflation settings.

Could not reach target pressure.

Check for air leaks.

Incorrect NIBP cuff size. Check
patient type.

The patient is outside of the module's
capability to measure NIBP.

Retake measurement.

Inflation too quick. Check NIBP cuff
and tubing connections.

Inflation was too quick; check for a
kinked hose.

Take corrective action; retake
measurement.

Unable to determine NIBP. Check
inflation settings.

Reading couldn’t be obtained
because a high enough inflation
couldn’t be reached.

Retake measurement.