1

Warnings and Cautions

Symbols

Caution: Consult user's manual for
additional information.

Warning: High temperatures

Risk of fire. Replace fuses as
marked.

Warning: High-intensity light

Warning: Power supply of unit is
energized whenever power cord is
plugged in.

Dangerous Voltage

UL Recognized Component

UL Listed

General Precautions

For safety, the colposcope should only be coupled to
a grounded 110-120 VAC hospital-grade outlet.

The colposcope should not be operated in the pres-
ence of flammable or explosive gases or chemicals,
or installed in areas where these materials are com-
monly used.

To avoid overheating, the colposcope should not be
positioned closer than 6” to any wall.

Keep all liquids away from electrical equipment to
avoid the possibility of shock and instrument dam-
age.

Occasionally inspect the power cord for signs of
cuts, abrasions or dents.

The colposcope should never be stored or operated
in areas where it could get wet or could be exposed
to any environmental conditions like extreme tem-
perature or humidity, direct sunlight, dust, etc.

The lamp is extremely bright. DO NOT stare directly
into illumination lens when the lamp is lit.

All service to the video colposcope must be per-
formed by Welch Allyn or by an authorized repair cen-
ter.

Colposcope user should adhere to the operating con-
ditions found in this manual. Otherwise, instrument
damage may occur and/or operator/patient safety
may be compromised.

There are no user serviceable parts (other than the
lamp and fuses) in this unit or in its accessories. Any
attempt to disassemble and/or repair this unit will
result in voiding of the warranty.

Colposcope device to patient distance no closer than
19 cm. Closer distances have higher irradiance levels
and may cause thermal skin injury. Short intermittent
exposures in and out of the illumination path by the
user at closer distances is typical of cervical exami-
nations. Prolonged exposures are not recom-
mended.

WARNING: DO NOT use a converter
adapter that will convert the three-prong
AC plug to a two-prong line plug. The
power supply in the colposcope will not
be properly grounded and electric shock
might result.

WARNING: The lamp operates at a high
temperature. DO NOT attempt to remove
the lamp before allowing it to cool. Allow
at least five minutes for the lamp to cool
before replacing.Replace with Welch Allyn
lamp #08100 only.

CAUTION: The device is for OB/GYN pur-
poses only.

CAUTION: Federal law restricts this
device to sale by on the or order of a physi-
cian or other appropriately licensed medi-
cal professional.