Classification I Type BF protected against defibrillator effects

Floating output circuit

Cautions, consult this manual for safety precautions

Pedal connection

Active handle connection

Patient plate connection

High voltage

ELECTRICAL

INPUT VOLTAGE:

95 - 135 VAC 50/60 Hz

190 - 250 VAC 50 Hz

CURRENT:

2.3 amps max.

2.3 amps max.

LOW FREQUENCY LEAKAGE:

Less than 50 micro-amps

Less than 50 micro-amps

FUSES:

2.5 amps, T type

1.6 amps, T type

POWER CORD:

American Cord Set, Hospital Grade,
NEMA 5-15P/IEC 320 Female, 250cm Long

PHYSICAL

Dimensions: 305mm x 267mm x 115mm Weight:

7.250 Kg

ELECTROSURGICAL OUTPUT

RF Output Frequency:

450 kHz

RF OUTPUT POWER:

Volts p-p Max

Duty

Crest

(open circuit)

Cycle

Factor

CUT

0-100 watts RMS*

830

---

1.4

BLEND

0-100 watts RMS*

1200

60%

2.0

COAG

0-80 watts RMS*

3800

10%

5.5

NOTE: *stable to >800 ohms (calibration at 500 ohms)

RF ISOLATION: less than 150 milli-amps at 200 ohms

CLASSIFICATION . . . . . . . . .I-Type BF
OUTPUT CIRCUIT . . . . . . . .Floating output. Protected against the effects of the

defibrillator.

WORKING MODE . . . . . . . .Discontinuous maximum duty cycle: 10/30 sec.

COOLING . . . . . . . . . . . . . .Convection cooling without fan

CONTROL . . . . . . . . . . . . .Foot pedal operated (pneumatic) with audible signals

and mode lights

Electrosurgical Unit

• Instruction Manual

11