Welch Allyn Electrosurgical Unit - User Manual User Manual
Page 12

Classification I Type BF protected against defibrillator effects
Floating output circuit
Cautions, consult this manual for safety precautions
Pedal connection
Active handle connection
Patient plate connection
High voltage
ELECTRICAL
INPUT VOLTAGE:
95 - 135 VAC 50/60 Hz
190 - 250 VAC 50 Hz
CURRENT:
2.3 amps max.
2.3 amps max.
LOW FREQUENCY LEAKAGE:
Less than 50 micro-amps
Less than 50 micro-amps
FUSES:
2.5 amps, T type
1.6 amps, T type
POWER CORD:
American Cord Set, Hospital Grade,
NEMA 5-15P/IEC 320 Female, 250cm Long
PHYSICAL
Dimensions: 305mm x 267mm x 115mm Weight:
7.250 Kg
ELECTROSURGICAL OUTPUT
RF Output Frequency:
450 kHz
RF OUTPUT POWER:
Volts p-p Max
Duty
Crest
(open circuit)
Cycle
Factor
CUT
0-100 watts RMS*
830
---
1.4
BLEND
0-100 watts RMS*
1200
60%
2.0
COAG
0-80 watts RMS*
3800
10%
5.5
NOTE: *stable to >800 ohms (calibration at 500 ohms)
RF ISOLATION: less than 150 milli-amps at 200 ohms
CLASSIFICATION . . . . . . . . .I-Type BF
OUTPUT CIRCUIT . . . . . . . .Floating output. Protected against the effects of the
defibrillator.
WORKING MODE . . . . . . . .Discontinuous maximum duty cycle: 10/30 sec.
COOLING . . . . . . . . . . . . . .Convection cooling without fan
CONTROL . . . . . . . . . . . . .Foot pedal operated (pneumatic) with audible signals
and mode lights
Electrosurgical Unit
• Instruction Manual
11
II.
SYSTEM FEATURES (cont.)