Technical specifications, Cardioperfect, Pro ecg recorder - user manual – Welch Allyn Cardioperfect PRO ECG Recorder - User Manual User Manual
Page 21

CardioPerfect
PRO ECG Recorder - User Manual
DIR 80012763 Ver.
F
21 / 26
4. Technical Specifications
Table 2 CardioPerfect Pro Cardiograph
PRO CARDIOGRAPH
SPECIFICATION
SE-PRO-600
Signal bandwidth
Sampling Rate
0.05 to 150Hz
600 samples/sec
SE-PRO-1200
Signal bandwidth
Sampling Rate
0.05 to 250Hz
1200 samples/sec
Pacemaker Bandwidth
2KHz
Amplitude Quantization
3.75 µV
Skew
72.9uSec
Input impedance
>5MOhms
Gain accuracy
5% RTI
DC dynamic span
±300 mV (95% gain accuracy)
AC dynamic span
±5mV
CMMR
83 dB
Recovery time after defibrillation
80% gain accuracy after 5s
Setup time after power-on
80% gain accuracy after 5s (all leads connected)
Power consumption
<750mW
Enclosure protection against ingress of water
None - IPX0
Degree of protection against electrical shock
BF
Table 3 CardioPerfect Pro Battery Pack
PRO BATTERY PACK
SPECIFICATION
Battery output voltage
2.85V - 4.5V
Battery type
Nickel Metal Hydride (NIMH)
Maximum recommended output current
700mA
Battery Capacity
1650mAh nominal
Lifetime battery charge cycles
500 charge-discharge cycles
Supply voltage
9
– 18 VDC
Charger supply rating
4W
Operating temperature
+10˚C to +40˚C, 30% to 70% RH
Charging temperature
0
˚C - 35˚C
Storage temperature
-
20˚C to + 40˚C
Table 4 Safety, EMC, and regulatory compliance
STANDARD
STANDARD
ANSI/AAMI EC11
UL60601-1
CAN/CSA C22.2 No. 601.1
IEC/EN 60601-1
CAN/CSA C22.2 No. 601.1.1
IEC/EN 60601-1-1
CAN/CSA C22.2 No. 601.1.2
IEC/EN 60601-1-2
CAN/CSA C22.2 No. 601.1.4
IEC/EN 60601-1-4
CAN/CSA C22.2 No. 601.2.25
IEC/EN 60601-2-25
IEC/EN 60601-2-51: 2003