Mechanical, Electrical, Environmental – Welch Allyn Spot Vital Signs - User Manual User Manual
Page 54: Mechanical electrical environmental

46
Welch Allyn Spot Vital Signs
Mechanical
Electrical
Environmental
Dimensions
Height: 9.7 inches (24.6 cm)
Length: 5.7 inches (14.5 cm)
Depth: 4.7 inches (12.0 cm)
Weight
Approximately 4.6 pounds (2.2 kg)
Mounting
Self-supporting on rubber feet
Custom Mobile Stand
Custom Wall Mount
Custom IV Pole Mount
Portability
May be hand-carried when held by the rear handle.
Power Requirements
Patient-rated isolation transformer is connected to AC mains:
North American Version: 120VAC, 60Hz. 0.13A Input, 7.2VDC, 1.0A Output
International Version: 230VAC, 60HZ, 0.065A Input, 7.1VDC, 0.860A Output
Australian Version: 240VAC, 50Hz, 13VA Input, 7.2VDC, 1.0A Output
Battery
Lead acid, with external charger.
A fully charged battery supports 130 typical blood pressure determinations taken at
7-minute intervals. The battery is 90-100% charged after 12 hours of charging. The
battery automatically charges when the Spot Vital Signs is powered through the AC
power transformer. The battery charges faster when the instrument is not in
operation.
Operating Temperature
+10
to +40C (Thermometer operating temperature 16° to 40° C)
+50
to +104F (Thermometer operating temperature 61° to 104° F)
Storage Temperature
-20
to +50C
-4
to +122F
Transport Temperature
-20
to +49C
-4
to +122F
Relative Humidity
15 to 90% (non-condensing)
Operating Altitude
-170 to +4877 m
-557 to +16,000 ft.