Welch Allyn Spot Vital Signs

Table 3. Temperature

Code

Description

Corrective Action

C20

Broken/missing probe

Replace probe.

P

Loss of tissue contact

Ensure proper probe positioning.

E0.2, E0.3

Ambient temperature out of range

Adjust ambient temperature or device location.

C22

10-minute diagnostic time exceeded

Remove probe, discard probe cover, retake
temperature.

Table 4. SpO

2

Code

Description

Corrective Action

E7

Internal SpO

2

error.

Retake reading.

C6

SpO

2

pulse rate out of range

Check patient condition.

C8

Faulty SpO

2

sensor.

Replace sensor.

C9

SpO

2

time limit exceeded.

Remove sensor from patient. Reapply sensor and
retake reading.