Specifications, Patient population, Blood pressure – Welch Allyn Spot Vital Signs - User Manual User Manual
Page 51: Temperature, 9 - specifications, Patient population blood pressure temperature, Blood pressure temperature

9
43
Specifications
Patient Population
The Welch Allyn Spot Vital Signs is designed for use with adult and pediatric patients.
Welch Allyn defines a pediatric patient as 29 days or more of age.
Welch Allyn defines neonates as children 28 days or less of age, born at term (37 weeks
gestation or more), otherwise up to 44 gestational weeks.
Blood Pressure
Temperature
WARNING THE WELCH ALLYN SPOT VITAL SIGNS IS NOT INTENDED FOR
USE ON NEONATAL PATIENTS.
Cuff Pressure Range
0 mmHg to 300 mmHg
Cuff Inflation Factory
Default
160 mmHg
Systolic Range
60 mmHg to 250 mmHg
Diastolic Range
30 mmHg to 160 mmHg
Accuracy
Blood pressure accuracy meets or exceeds SP10-1992 AAMI standards for
non-invasive blood pressure accuracy (AAMI standard: ± 5 mmHg mean error,
8 mmHg standard deviation). Blood pressure accuracy is validated for pressure
measurement using the upper arm only.
Determination Time
Typical: 20 to 45 seconds
Maximum: 165 seconds
Pulse Rate Range
40 bpm to 200 bpm
Pulse Rate Accuracy
±5.0%
Overpressure Cutoff
305 mmHg -0/+15 mmHg
Accuracy
Calibration accuracy: + 0.2° F (+ 0.1° C).
Range
Maximum: 109.4° F/43.0° C
Minimum:
86.0° F/30.0° C
Determination Time
Oral: approximately 4 seconds
Axillary: approximately 10 seconds
Rectal: approximately 15 seconds