Table 12. Blood Pressure Cuff Pops Off

Possible Cause

Explanation and Corrective Action

Inappropriate blood pressure
cuff size

Determine blood pressure cuff size with the blood pressure cuff markings or see
“Chart for Determining Blood Pressure Cuff Size” on page 23.
If blood pressure cuff
continues to pop off, notify biomedical department or Welch Allyn Technical Support.

Blood pressure cuff applied
inside out

Re-apply blood pressure cuff. Make sure Welch Allyn label is facing away from arm.

Table 13. Blood Pressure Cuff Deflating Too Slowly

Possible Cause

Explanation and Corrective Action

Normal operation

Typical time to take a reading is 20 to 45 seconds; 165 seconds is the maximum.

Pressure preset too high

Check default pressure preset setting in internal configuration mode.

Patient movement

Have patient sit still. Do not have arm tight against chest wall, as respiration may
affect speed and accuracy of blood pressure measurement.

Small leak in pneumatic
system

Check blood pressure cuff tubing and pressure hose for leaks.