Standards, Specifications, and
Accessories

Standards and Compliance

The AM282 meets

Frequency Range

Intensity

*The above intensity ranges apply to the external headset earphones.

ANSI S3.6 (2004)

Specification for Audiometers (Type 4)

IEC 60645-1 (2002)

Electroacoustics - Audiological Equipment - Pure Tone Audiometers (Type 4)

IEC 60601-1 (2003)

Medical Electrical Equipment Part 1 - General Requirements for Safety

UL 60601-1

Medical Electrical Equipment, Part 1 -General Requirements for Safety

CAN/CSA-C22.2 No. 601.1-M90 (2003)

Medical Electrical Equipment - Part 1: General Requirements for Safety

IEC/EN 60601-1-2 (2001)

Medical Electrical Equipment Part 1 - Electromagnetic compatibility

requirements

Discrete Frequencies:

125, 250, 500, 750, 1000, 1500, 2000, 3000, 4000, 6000, 8000 Hz

Accuracy:

±2%

Total Harmonic Distortion:

<2.5% measured acoustically at the maximum Hz for frequencies below

5000 Hz and measured electrically above 5000 Hz.

Ranges* In increments of 5 dB steps

125 Hz

-10 to 50 dB HL

500 to 4000 Hz

-10 to 90 dB HL

6000 Hz

-10 to 85 dB HL

250 and 8000 Hz

-10 to 70 dB HL

Note

The maximum output values in dB HL are reduced by 10 dB when insert phones
are used, except for 6 kHz where the maximum dB HL is reduced by 20 dB.

A “+10” dB button extends the maximum at all frequencies by 10 dB.