beautypg.com

Testing methods, Unlocking methods – Rice Lake MA160 Sartorius Moisture Analyzer User Manual

Page 48

background image

48

Moisture Analysis

Operating Instructions Moisture Analyzer MA160

Testing Methods

Evaluation

Result is within the tolerance range
Result: 85.15 %M
Target value: 85.00 %M
Tolerance range: 84.50 %M - 85.50 %M
Are you satisfied with this result?

NO

YES

y If the target value and initial weight options are activated, the measurement/

method is evaluated and the result of the evaluation is displayed .

t

Select the desired option:

– If you are satisfied with the result, select

YES

. You can then unlock the

method .

– If you are not satisfied with the result: Select

NO

and run the test again .

Test the method

Method successfully created. Would you

like to test the method now?

Temperature: 105 °C

NO

YES

t

Select the desired option:

– To test the method with the suggested temperature: Select

YES

. Moisture

analysis is then carried out with the set parameters (see „Performing the
Moisture Analysis“ on page 38)
.

– To save the method without testing: Select

NO

and then

o

to return to the

method menu .

Unlocking Methods

+

Standard

Coffee creamer

100 °C
Fully automatic

Oats

80 °C
Fully automatic

Wood glue

110 °C
Fully automatic

105 °C
Fully automatic

Ketchup

110 °C
Fully automatic

Methods must be unlocked (available) before they can be used .

Unavailable (locked) methods are marked in the method menu with an

H

symbol .

t

Select the method you wish to unlock .

Customize parameters

Heating program

End

Initial weight

Result display

Standard drying
110 °C

Fully automatic

Off

Moisture in %M

Method name

Ketchup

y The parameters of the method are displayed .

Customize parameters

Test

Release

Delete

Print

t

Scroll all the way to the bottom .

t

Select

Release

.