Timeout standby temp, Usb port – Rice Lake MA160 Sartorius Moisture Analyzer User Manual
Page 31

31
System Settings (Setup Menu)
Operating Instructions Moisture Analyzer MA160
Timeout standby temp.
Timeout Standby Temp.
2
1
3
Time [h]
The timeout for the standby temperature can be set here .
t
To set the time, select
t
, enter the desired time and confirm with
l
(entry in
hours from 1 h to 24 h) .
USB port
USB port
Device / Protocol
RS232 Configuration
Decimal separator
Keyboard emulation
The settings for the data transfer to a peripheral device (such as a PC or printer)
can be defined here .
Depending on the cable and peripheral device connected to the moisture analyzer,
various options are available in the menu . The moisture analyzer automatically
detects the type of connection .
There are several ways to connect the device via USB:
1) Connection to a PC via USB Cable (Mini-B to USB A)
Device / Protocol
PC - SBI
PC spreadsheet format
PC text format
Off
Under
Device/Protocol
:
t
Select the data format for the transfer to the PC .
–
PC - SBI
: Driver for PC required (CDC Virtual Com Port) . The data is sent
via SBI protocol using a virtual serial interface . Additional information can
be found in Chapter “USB Interface”, section “Direct Transfer of Data (PC)” .
–
PC spreadsheet format
: The analyzer transmits the data via keyboard
command (keyboard emulation) to the currently opened application on the
PC in spreadsheet format .
–
PC text format
: The analyzer transmits the data via keyboard command
(keyboard emulation) to the currently opened application on the PC in text
format .
–
Off
: Data transmission is deactivated .
Mini-B
PC-SBI
PC spreadsheet format
PC text format
Off
Printer YDP20-0CE
USB-RS 232 SBI
Off
Printer YDP30,
YDP40
Off
Type-A
Mini-A
Type-B
Mini-A
RS 232
9-pin
Sartorius order no.:
YCC04-D09
Sartorius order no.:
YCC03-D09
USB Host
(Master)
1)
2)
3)
USB Device
(Slave)
Serial device
with 9-pin
SUB-D plug
Printer equipment
supplied