0 calibration, 1 front panel calibration – Rice Lake IQ Plus 210 Digital Weight Indicator User Manual
Page 24

20
IQ plus 2100 Installation Manual
4.0
Calibration
The IQ plus 210 can be calibrated using the front panel, EDP commands, or the Revolution
™
configuration utility.
Each method consists of the following steps:
•
Zero calibration
•
Entering the test weight value
•
Span calibration
•
Optional rezero calibration for test weights using hooks or chains.
The following sections describe the calibration procedure for each of the calibration methods.
Figure 4-1. Calibration (CALIBR) Menu
4.1
Front Panel Calibration
To calibrate the indicator using the front panel, do the
following:
1. Place the indicator in setup mode (display
reads
CONFIG
) and remove all weight from
the scale platform. If your test weights require
hooks or chains, place the hooks or chains on
the scale for zero calibration.
2. Press the
RIGHT
(
UNITS
) button until the
display reads
CALIBR
ENTER
(
ZERO
) to go to zero calibration
(
WZERO
).
3. With
WZERO
displayed, press
ENTER
to
calibrate zero. The indicator displays
°CAL°
while calibration is in progress. When
complete, the A/D count for the zero
calibration is displayed. Press
ENTER
again to
save the zero calibration value and return to the
WZERO
prompt or use the procedure shown in
Figure 4-2 to edit the value. When done, press
RIGHT
to go to the
WVAL
prompt.
4. With
WVAL
displayed, press
ENTER
to show
the test weight value. If the value is equal to the
test weight you are using, press
ENTER
again
to save the value and return to the
WVAL
prompt; if the value is incorrect, use the
procedure shown in Figure 4-2 to edit the
value. When done, press
RIGHT
to go to the
WSPAN
prompt.
NOTE:
When calibrating the indicator with
LB/OZ as the primary unit, enter the WVAL value
in ounces. For example, to calibrate a 5 lb scale
using LB/OZ, enter 80 (5 lb x 16 oz/lb) as the
WVAL value.
5. With
WSPAN
displayed, place test weights on
the scale then press
ENTER
to calibrate span.
The indicator displays
°CAL°
while calibration
is in progress. When complete, the A/D count
for the span calibration is displayed. Press
ENTER
again to save the span calibration value
and return to the
WSPAN
prompt or use the
procedure shown in Figure 4-2 to edit the
value. When done, press
RIGHT
to go to the
REZERO
prompt.
Figure 4-2. Editing Procedure for Numeric Values
WZERO
*CAL*
Display and edit
zero calibration
A/D count value
Display and edit
test weight value
WVAL
*CAL*
Display and edit
span calibration
A/D count value
WSPAN
*
CAL*
Press Enter to
remove offset from
zero and span
calibrations
REZERO
XXXXXXX
XXXXXXX
VERS
DEFLT
DIG IN
SERIAL
CALIBR
CONFIG
FORMAT
To edit numeric values, press the
UNITS (RIGHT)
button to select the leftmost digit. Each digit flashes
when selected: Press
ZERO (ENTER)
to
increment the value of the selected digit; press the
RIGHT
button to move right to the next digit.
Once the last digit has been edited, press
RIGHT
again and the digit stops flashing. Press
ENTER
to save the value entered and return to the menu level
above, or press
RIGHT
to edit the value again,
starting with the leftmost digit.
0
0 0 0 0 0