Rice Lake Lift Truck/Pallet Jack Scales - Intercomp PW800 User Manual
Page 14

PW800, Users
Rev O, January, 2008
14
14. The scale shows “EE-06
EE-06
EE-06
EE-06”. Press the NET/GROSS and MODE keys. Enter demand
versus continuous on the optional serial output.
See Serial Out section
Setting
Type
0
Continuous
1
Demand
Check for calibration blocking switch.
At this point the scale saves any changes that have been made. A check is than made to see
whether or not the calibration blocking switch is enabled (CAL). If enabled the calibration
procedure is continued; if disabled (RUN) the scale returns to normal weighing.
Last seven parameters.
15. The scale shows “EE-07
EE-07
EE-07
EE-07”. Press the NET/GROSS and MODE keys. Enter the AZT
size (0 to 3). The AZT size is the number of graduations the auto zero tracking can
remove.
Setting
AZT size
0
Off
1
0.6
2
1
3
3
16. The scale shows “EE-08
EE-08
EE-08
EE-08”. Press the NET/GROSS and MODE keys. Enter whether
the zero range is on or off (0 to 1). If the zero range is ON the zero may move no
more than 5% away from the original zero obtained at calibration.
Setting
Zero range
0
Off
1
On
17. The scale shows “EE-09
EE-09
EE-09
EE-09”. Press the NET/GROSS and MODE keys. The scale
shows the current Canadian specification selection. Use the following table to select
the Canadian specifications setting. When Canadian specifications are set (1):
EE-08
EE-08
EE-08
EE-08, EE-09
EE-09
EE-09
EE-09, and EE-11
EE-11
EE-11
EE-11 have no meaning.
Setting
Canadian
Specification
0
Off
1
On
A '0' setting implies normal operation:
1: AZT size is determined by the setting of EE-10
EE-10
EE-10
EE-10.
2: Zero operates over full range allowed by EE-11
EE-11
EE-11
EE-11.
3: The over-capacity point is determined by the setting of EE-13
EE-13
EE-13
EE-13.
A '1' setting implies Canadian specifications are used.
1: The AZT size fixed at 0.6d regardless of EE-08
EE-08
EE-08
EE-08 setting.