0 troubleshooting, Troubleshooting – Rice Lake Equipoise Series, Rice Lake Weighing Systems User Manual
Page 21

Replace Battery 17
8.0
Troubleshooting
Problems
Possible cause
Common Solutions
Display is
blank.
No self test
Main power is turned off.
Power supply faulty or not
plugged in.
Internal battery is not charged.
On/Off switch is faulty.
Check power is getting inside the
balance and on/off switch is working.
Verify the voltages, which are on the
power labels.
Blank display
after self test
Pan not installed
Unstable weight
Load cell damaged
Check the pans are installed
correctly.
Try to turning on again.
OL or -------
Maximum capacity exceeded
Load cell or mechanics
damaged
Power supply faulty
Check the platform is installed
correctly.
Try to turn on the balance again.
Do the calibration again
------- or
NULL
Weight is on the platform is
below permissible limit.
Pan not installed correctly.
Power supply faulty.
Load cell or mechanism faulty
Check the platform is installed
correctly.
Try to turn on the balance again.
Do the calibration again
Display is
unstable
Goods touching somewhere.
Air variation or any vibrations.
Temperature changed .
Load cell or connections faulty.
Power supply faulty
Check the balance is in acceptable
location.
Check the connecters and load cell.
Check the power supply and battery
Weight value
incorrect
Calibration error.
Platform of load cell touching
somewhere.
Wrong weighing unit
Use accurate weight for to do the
calibration
Check the pan and load cell is
installed proper and touching.
Check the parameter settings.
Check the load cell and connecters
Can not use
full capacity
Over load protection stoppers or
transport locks are not removed.
Parameters are set incorrectly.
AD problem.
Load cell or mechanism
damaged
Check the stoppers and locks under
the platform.
Check the weighing unit and
parameter settings.
Check the load cell.
Platform
Corner
Weight
different
Over load protection stoppers or
transport locks are not removed.
Load cell or mechanism
damaged
Check the stoppers and locks under
the platform.
Use accurate weight for to do the
calibration
Check the load cell.
Battery not
charging
Mains voltage problem
Charging circuit problem
Battery Problem
Check the mains and adaptor.
Check the battery.
Check the charging circuit