The results from the flow meter tests indicate what maintenance needs to be
performed on the air pump. Refer to the following table to evaluate the test results. If
the flow meter reading is equal to or lower than the dirty air pump reading, follow the
steps outlined in the action required column of the table. Pump head rebuild kits (P/N
79190003) are available from Dynojet. Refer to the Air Fuel Ratio Module Installation
Guide (P/N 98295110) for rebuild instructions.
Note: Before replacing the pump head service cover be sure to perform any
necessary maintenance procedures.
test point
new or rebuilt
pump
dirty pump
action required
test point one
35
25 or less
rebuild pump heads
test point two
28
25 or less
clean or replace filter
test point two—filter
assembly
less than 2
greater than 2
clean and tighten filter
housing
test point three
28
25 or less
rebuild pump heads
test point three—sensor
block assembly
less than 2
greater than 2
clean and tighten sensor
test point four
30
25 or less
rebuild pump heads
test point five
30
25 or less
rebuild pump heads