27

a.

Blown fuse.

a. Check for electrical short; replace

fuse.

d.

Transmitter damaged or faulty.

d. Replace transmitter indicated by

controller.

6. Controller error text, “Transmitter

Error”.

c. Wiring

fault.

c.

Check

and fix transmitter wiring.

a.

Bulk liquid withdrawal valve
closed.

a. Open

valve.

d.

Bulk liquid withdrawal line
strainer clogged.

d. Clean strainer.

c. Vent to bulk solenoid valve (V-

2) failure.

c. Check fuse and wiring. Rebuild or

replace solenoid valve if
necessary.

7. Controller error text, “Vent to

Bulk Time Out”.

d. Pressure transmitter (PT-2) or

bulk tank pressure transmitter
malfunctioning.

d. Check against accurate pressure

gauge. Replace if necessary.

a.

Bulk liquid withdrawal valve
closed.

a. Open

valve.

d.

Bulk liquid withdrawal line
strainer clogged.

d. Clean strainer.

c. Bulk tank empty or pressure low. c. Refill or repair bulk tank.
d. Vent solenoid valve (V-3)

failure.

d. Check fuse and wiring. Rebuild or

replace solenoid valve if
necessary.

8. Controller error text, “Fill Time

Out”.

e. Level transmitter (PT-1) requires

calibration or is malfunctioning.

e. Calibrate level transmitter per

“First Fill” section of manual.
Replace level transmitter if
necessary.

a.

Bulk liquid withdrawal valve
closed.

a. Open

valve.

d.

Bulk liquid withdrawal line
strainer clogged.

d. Clean strainer.

9. Controller error text, “Filling

Error”.

c. Excessive product withdrawal.

c. Check for leaks downstream.

Reduce product use.