Scilogex DM1424 Hematocrit-Centrifuge User Manual
Page 20

17
11.1.3 Available rotors and adapters
All the rotors are made of plastic, they cannot be high-pressure sterilization and UV irradiation, only
ordinary sterilization can be used.
Rotor
type
ID
code
Tubes
Adapters
Maximum
speed
(rpm)
Maximum
centrifugal
radius
r
max
(cm)
Maximum RCF
Rcf (×g)
1 A24-2P
24-2/1.5ml tube
14000
8.5
18620
24-0.2ml PCR
tube
A02P2
14000 6.9 15110
24-0.5ml micro
tube
A05P2 14000
7.6 16638
2 AC24 24-ш1.75Ч75 12000
8.5
13680
Table 11.1 Rotors and adapters
11.1.3 Notice
(1) The centrifuge rotor can separate samples with a density lower than 2.0g/ml. If the samples density is
over 2.0g/ml, please calculate allowable speed depending on the following formula.
Allow Speed (rpm)= Maximum speed×(2.0(g/ml)/Sample density (g/ml))
1/2
(2) To prevent corrosion, remove the rotor from rotor chamber if not in use for a long term, then detach the
rotor lid and place upside down to dry the tube holes.
(3) If samples have leaked in the rotor holes, wash the hole with water, apply a thin coat of silicon grease
on the rotor surface after drying.
(4) It is necessary for a regular rotor maintenance and should be cleaned every 3 months to keep the tube
holes and shaft clean. Apply a thin coat of silicon grease.