10.2 How to open the door
10.2.1 In the case of
power on
(
1) Turn on the power switch, release the door automatically.
(
2) The door will be released automatically once the operation is finished.
(
3) It is available to release the door by press
button once the rotor stops.
10.2.2 In the case of power outage
The door cannot be opened automatically if there is a power outage. It is available to be opened manually
as follows.
(
1)Ensure if the rotor has stopped rotating.
z Listen carefully to ensure no rotating sound can be heard.
(
2)Insert a screw driver into a hole to open door.
z Holes are located on the left and right sides of the unit.
z Insert a screw driver into the two holes and push forward to release the door.
10.3 Replacement of fuses
(
1)There are two fuses, 250V, 5A time-delay type, size: Ф5×20. There are two fuses of D3024R, 250V,
10A time-delay type, size: Ф5×20.
(
2)The fuse holder is located in the power inlet. Pull out the fuse holder from power inlet and replace the
fuses if necessary.
11. Instructions for the rotor and tubes
CAUTION
z Read the instructions thoroughly, to properly load and use rotor.
z Do not exceed the allowable maximum speed of rotor、tube and adapters etc. Ensure the allowable
maximum speed of adapters is lower than the rotor’s maximum speed.
CAUTION
z The door just can be opened while the power is on and rotor stops rotating.