Rotor preparation – Scilogex DM1424 Hematocrit-Centrifuge User Manual

Page 10

background image

7

Figure 6-2 The main interface

Main interface is as figure 6-2. The speed is set to be 14000rpm, temperature of

centrifugal chamber is

25℃, and the running time is 12 minutes. When speed symbol

is rotating, this indicates the centrifuge is

running. If the rotation is faster, the speed is higher. Temperature of chamber is displayed and cannot be
controlled. Time symbol

displays the ratio of working to time setting. The total time setting is divided

into 10 sections.

7. Rotor Preparation

7.1 Prepare the samples

7.2 Inject the samples into tubes

z Do not exceed the actual capacity allowed in the user manual.

7.3 Keep the tubes balanced

z Although the centrifuge can accept sample balancing by eye, we recommend that you keep this

centrifuge in a well-balanced condition to extend its life expectancy.

z Never intentionally run the centrifuge under an unbalanced condition even though the allowable

imbalance is not exceeded.

7.4 Inspect the rotor

Check the rotor for corrosion or scratches before using.

CAUTION

z If any abnormality such as corrosion or scratches are found, stop using the rotor and contact our

service center.

z Only manufacturer’s rotors must be used with the unit.

CAUTION

z Do not overload samples into the centrifuge which will cause leaking.

14000 25

12

min

rpm