8.3 Pulse operation
This function is used to remove the residual samples adhered to the interior of the tubes or for quick spins.
(
1)Turn on the power switch and load the rotor to the shaft, tighten the rotor lid and make sure it is in
secured position, and then close the door.
(
2)The centrifuge goes into preparation mode and displays last running values.
(
3)Press
knob and hold, the centrifuge will speed up to the setting speed. While releasing the
knob during acceleration, the centrifuge will start to decelerate and stop.
9. Maintenance
9.1 Cleaning
CAUTION
z If do not follow the recommended instructions for cleaning or disinfecting this may damage the
centrifuge.
(
1)Centrifuge
z If the centrifuge is exposed to ultraviolet rays for a long time, the color of the door may be changed or
the label may be peel off. After using, cover the centrifuge with a piece of cloth to protect it from
direct exposure.
z If the centrifuge needs cleaning, clean it with a cloth or sponge moistened with a neutral detergent
solution.
z Sterilize the centrifuge by wiping with a cloth moistened with 70% ethanol solution.
(
2)Rotor chamber
z If the rotor needs cleaning, clean with cloth or sponge moistened with a neutral detergent solution.
Sterilize the centrifuge by wiping with a cloth moistened with 70% ethanol solution.
Note:The button works only while the rotor stopped and the door is locked.
CAUTION
z Do not directly pour water, neutral detergent or disinfectant solution into the rotor chamber, otherwise
fluids may leak into the drive units and cause corrosion or deterioration to the bearings.