10. Troubleshooting

10.1 Possible problems and solutions

This centrifuge has a self-diagnostic function. If a problem occurs, an error/warning code will be displayed

on the time display screen, and the operator can determine the malfunction with the alarm code below

.

Symptom Causes

Solutions

Nothing appears on the screen when the

power is turned on.

·Facility power circuit

breaker tripped.

·The fuse has blown.

·Correct and turn on the

power.

·Replace the fuse.

Error code appeared on the

time display screen

E-02

Door fault

·The door opened while

running.

·Press the button

with the door opened.

·Close the door

immediately.

·Close the door,and then

start to operate.

E03

Rotor identify

fault

The centrifuge cannot

identify the rotor.

·Use the correct rotor.

·Replace rotor identify

connection.

E-04

Temperature

fault

·The air vents are blocked.

·Radiator fan is damaged.

·Clean air vents.

·Replace radiator fan.

E-06

Set wrong

value

·The setting value exceeds

the allowable range.

·Modify the set value.

E-10~86

·Read the service manual.

·Contact the service center.

Table 10-1 Possible problems and solutions

z

Alarm codes E-1~E-9 are related to incorrect operation/programming. You can continue running the

centrifuge after implementing corrective procedures.