Scilogex DM1424 Hematocrit-Centrifuge User Manual
Page 12

9
Figure 8-2 indicates the centrifuge has accumulated running time 312 hours 56 minutes and 45 seconds,
and then the centrifuge displays the last running parameters, see figure 8-3 below:
Figure
8-3
Last
running interface
z
Speed: 14000rpm; running time: 12minutes; centrifugal chamber temperature: 25℃
z Release the door.
8.1.1 Load and replace the rotor
Load the rotor to the shaft Locking nut Load the rotor lid
Figure 8-4 Load the rotor
CAUTION
z Attach the rotor to the rotor shaft. Ensure the rotor is in position and connected with the shaft,
tightening the locking nut to secure the rotor with shaft, to prevent the rotor damaging the centrifuge.
z Ensure the rotor lid is firmly tightened to the rotor.
z Load the rotor to the shaft to ensure rotor is in position until it connected with the shaft.
z You should feel a ‘click’ when the rotor is properly loaded on the shaft. If not, there may be something
stuck between the rotor and the shaft. Double check and clean it if necessary.
z Rotate the rotor slightly with your fingers to check if the rotor vibrates, if so attach the rotor again.
z Rotate the nut clockwise using the wrench to tighten the rotor to the shaft firmly.
z Close the rotor lid, firmly tighten clockwise the lid to the rotor and ensure is in position. Close the
door and then start running.
z The method of removing the rotor is as same as the above mentioned by turning the locking nut
counterclockwise.