2 rcf operation – Scilogex DM1424 Hematocrit-Centrifuge User Manual
Page 14

11
(
3)Warning display
z If an error occurs during the operation, the centrifuge will brake to stop automatically, and display the
error code on the time/display area. The error code can be checked in the table 10-1, and corrective
actions can be applied accordingly.
8.1.4 End the operation
(
1)The centrifuge will brake when it reaches the set time or
button is pressed.
z When the rotor stops rotating, the centrifuge will start beeping to alert the operation has finished.
(
2)Open the door
z The door can be released automatically when the operation has stopped.
z With the door closed, you are able to press the
button to open it.
z After ending the operation, the program will store the setting parameters of this operation, and will
recall these parameters when restarting the program.
(
3)Open the door and take out the rotor and samples.
8.2 RCF operation
(
1) Turn on the power switch
(
2) Set a RCF (Relative Centrifugal Force) value
CAUTION
z Do not exceed the allowable maximum RCF value of the rotor and adapters.
z Press the select button
and choose speed unit ×g, the speed symbol will flash into RCF value
input status.
z If no button is pressed after the speed value has flashed after 5 seconds, the input mode will be shut
down.
z Rotate program button
to input a RCF value, RCF increment is10×g.
(
3)Set operating conditions
The other operation, please refer to the section 8.1.