8.1.2 Set the operation programs
Press the
button to select required program. The parameter can be modified when the program is
flashing. Rotate the program button clockwise
to increase parameter value. Rotate the program
button clockwise
to decrease parameter value. Rotate the program button
faster, and the
parameter value will increase faster. The minimum speed increment is 10 rpm, the minimum time
increment is 1 second.
(
1)Set the speed
z Press the select button
until the speed rpm is displayed.
z When the speed button is selected, the speed symbol will flash the speed value.
z The minimum speed value you can set 200rpm, the minimum increment is 10rpm.
z Rotate program button clockwise
to increase speed value. Rotate the program button
anti-clockwise
to decrease speed value.
z You can speed-up set the speed value by rotating program button
quickly.
z There is a circulating function to increase/decrease the speed values. Rotate the program button
clockwise
to change settings from small → large → maximum → minimum. Rotate the program
button anti-clockwise
to change settings from large → small → minimum → maximum.
(
2)Set the time
z Press select button
, time value flashes in the time setting mode.
z Rotate the program button
to set running time from 30 seconds to 99 minutes.
z When the time displays HD, this is a continuous running mode.
8.1.3 Start the operation
(
1)Press button
to start running
z The door must be locked before rotor starts spinning.
z The timer will start once the rotor starts spinning, the screen displays the remaining run time.
(
2)View and modify the operation programs
z Operation programs can be modified after the centrifuge reaches the set speed.
z Pressing the select button
, returns the display to the program interface and displays setting
programs. Press the select button
to the desired program. When flashing, rotate parameter
button
to modify values. Release the button after 5 seconds, and the centrifuge will return to
normal operation mode and run according to the new value.
z If the set time value has been modified, the operation time is not affected and will continue.