2 consumables, 3 routine inspection – Scilogex DM1424 Hematocrit-Centrifuge User Manual
Page 16
13
(
3)Drive shaft
z We recommend regular maintenance for drive shaft. You can wipe the drive shaft with soft cloth, and
then apply a thin coat of silicon grease.
(
4)Door
z Clean and sterilize the door using the same method as the section(1)above.
(
5)Rotor
z To prevent corrosion, remove the rotor from rotor chamber. If not in use for a long term, then detach
the rotor lid and turn upside down to dry the tube holes and keep clean.
z For sample leaks in the rotor, rinse the rotor with water. Apply a thin coat of silicon grease to the rotor
when it is completely dry.
z The rotor should be checked every 3 months to ensure the tube and rotor holes keep are clean and
apply a thin coat of silicon grease.
9.2 Consumables
Replaceable wearing parts listed below. It is recommended to replace these according this table.
Item
Replacement parts
Replacement conditions
1
Seal ring of door
Cracked
2
Rubber housing of temperature sensor
9.3 Routine inspection
(1) Check that the centrifuge is on a firm, flat and level surface, ensure the four feet are seated on the
surface firmly.
(2) Check that the centrifuge is grounded properly. Use a multimeter to check if there is a short circuit
between the power cord grounding pin and the motor shaft. If an open circuit is found, check for reasons
and troubleshoot before use.