2 consumables, 3 routine inspection – Scilogex DM1424 Hematocrit-Centrifuge User Manual

Page 16

background image

13

3)Drive shaft

z We recommend regular maintenance for drive shaft. You can wipe the drive shaft with soft cloth, and

then apply a thin coat of silicon grease.

4)Door

z Clean and sterilize the door using the same method as the section(1)above.

5)Rotor

z To prevent corrosion, remove the rotor from rotor chamber. If not in use for a long term, then detach

the rotor lid and turn upside down to dry the tube holes and keep clean.

z For sample leaks in the rotor, rinse the rotor with water. Apply a thin coat of silicon grease to the rotor

when it is completely dry.

z The rotor should be checked every 3 months to ensure the tube and rotor holes keep are clean and

apply a thin coat of silicon grease.

9.2 Consumables

Replaceable wearing parts listed below. It is recommended to replace these according this table.

Item

Replacement parts

Replacement conditions

1

Seal ring of door

Cracked

2

Rubber housing of temperature sensor

9.3 Routine inspection

(1) Check that the centrifuge is on a firm, flat and level surface, ensure the four feet are seated on the

surface firmly.

(2) Check that the centrifuge is grounded properly. Use a multimeter to check if there is a short circuit
between the power cord grounding pin and the motor shaft. If an open circuit is found, check for reasons
and troubleshoot before use.