Parameter commands – Baumer N 152 User Manual
Page 15
N 152
www.baumer.com
15
4.2.9. Indicate optional column of figures in upper line „t“ (74h)
This command is utilized to indicate a 5-digit number in the upper line of the display. The numerical column is
indicated less dot respectively comma. Preceding zeroes as well as both arrows are suppressed. The bottom
line still indicates the actual value.
The numerals are indicated in the display until any command except „t“, „u“ or „R“ is received. The number will
be retained after power failure.
Example: (Column of figures = 054321)
SOH
Adr
Cmd
Column of figures = 054321
EOT
CRC
Transmission
01h
20h
74h
30h
35h
34h
33h
32h
31h
04h
C6h
SOH
Adr
Cmd
Column of figures = 054321
EOT
CRC
Response
01h
20h
74h
30h
35h
34h
33h
32h
31h
04h
C6h
4.2.10. Indicate optional column of figures in bottom line „u“ (75h)
The command is utilized to indicate a 5-digit column of figures in the bottom line of the display. The number is
indicated less dot respectively comma. Preceding zeroes as well as both arrows are suppressed. The upper line
still indicates actual target or the column of figures.
The column is indicated until any optional command except „t“, „u“ or „R“ is received by the interface. The
number will be retained after power failure.
Example: (Column of figures = 012345)
SOH
Adr
Cmd
Column of figures = 012345
EOT
CRC
Transmission
01h
20h
75h
30h
31h
32h
33h
34h
35h
04h
B6h
SOH
Adr
Cmd
Column of figures = 012345
EOT
CRC
Response
01h
20h
75h
30h
31h
32h
33h
34h
35h
04h
B6h
Note:
Command “t” in conjunction with command “u” enables indication of a 10-digit number.
Both columns are not saved in EEPROM.
4.3. Parameter commands
Command
code
Data amount
in bytes
read
write
Broad-
cast
Saved in
EEPROM
Significance
a (61h)
5
X
X
-
X
Read / write general parameters
b (62h)
8
X
X
-
X
Parameter: tolerance compensation, window
c (63h)
8
X
X
-
X
Parameter: Scaling of spindle pitch
g (67h)
12
X
X
-
X
Parameter: Limit positions MIN, MAX
h (68h)
12
X
X
-
X
Parameter: Switching points motor speed
i (69h)
1
X
X
X
X
Parameter: mm / inch
j (69h)
3
X
X
X
X
Parameter: Timeout bus error
k (6Bh)
9
X
X
-
X
Parameter: motor system times
l (6Ch)
5
X
X
-
X
Parameter: Jog function
m (6Dh)
5
X
X
-
X
Read / write general parameters