that have flat and concave surfaces “on edge” will enable the
condensate to drain off and more readily dry, usually without
spotting. An additional cause of spotting can be traced to the
instrument wraps. During laundering procedures, it is vital that
detergents are thoroughly rinsed out, and that the final rinse is
prepared so that the wraps have a pH between 6.8 and 7.0. In
addition, healthcare professionals should check the cleanliness
of the sterilizer chamber. Steam can lift soil and poorly rinsed
chamber cleaning detergents from the chamber walls and deposit
them onto instruments and wrappers.
Rust Deposits: It is very unlikely for surgical grade steel to rust. What
appears to be rust is often residual organic matter in box locks or
mineral deposits which have been baked onto the surface of the
instrument. In localities where the water has a high iron content,
for example, an iron deposit will result in a metallic film on the
instrument. This may be prevented with the use of “treated water”
for the FINAL rinse during cleaning procedures.
The most effective method of dealing with instrument problems is
to prevent them from occurring. The use of “treated water”, careful
preliminary cleaning, using neutralized pH solutions, following
manufacturer’s instructions, and visual inspection, will help to
keep instruments performing accurately and cosmetically free of
troublesome stains. It is important to act quickly should a problem
arise. Delay will compound the problem and irreparable harm may
result.
RETURNED GOODS POLICY
Products must be returned in unopened packages with
manufacturer’s seals intact to be accepted for replacement or
credit unless returned due to a complaint of product defect.
Determination of a product defect will be made by Miltex. Products
will not be accepted for replacement if they have been in the
possession of the customer for more than 120 days.
REPAIRS AND MAINTENANCE
Should your instruments require repair or maintenance, contact
Miltex for return authorization and address. Instruments returned
to Miltex for repair must have a statement testifying that each
instrument has been thoroughly cleaned and sterilized. Failure
to supply evidence of cleaning and disinfection will result in a
cleaning charge and delayed processing of your instrument repair.
PRODUCT INFORMATION DISCLOSURE
MILTEX, INTEGRA AND MANUFACTURER EXCLUDE ALL
WARRANTIES, WHETHER EXPRESSED OR IMPLIED, INCLUDING
BUT NOT LIMITED TO, ANY IMPLIED WARRANTIES OF
MERCHANTABILITY OR FITNESS FOR A PARTICULAR PURPOSE.
NEITHER MILTEX, INTEGRA NOR MANUFACTURER SHALL BE
LIABLE FOR ANY INCIDENTAL OR CONSEQUENTIAL LOSS,
DAMAGE, OR EXPENSE, DIRECTLY OR INDIRECTLY ARISING
FROM USE OF THIS PRODUCT. NEITHER MILTEX, INTEGRA NOR
MANUFACTURER ASSUME NOR AUTHORIZE ANY PERSON TO
ASSUME FOR THEM ANY OTHER OR ADDITIONAL LIABILITY OR
RESPONSIBILITY IN CONNECTION WITH THESE PRODUCTS.
SYMBOLS USED ON LABELING
Manufacturer
Authorized Representative in the European Community
Catalog number
Lot Number
See instructions for use
Consult instructions for use
Product complies with requirements of directive 93/42/
EEC for medical devices
LOT
LOT
REF
Integra York PA, Inc.
589 Davies Drive
York, PA 17402
toll free phone: 866-854-8300
phone: 717-840-9335
fax: 717-840-9347
PAD-IFU-GEN Rev. B 06/13
Padgett is a registered trademark of Integra LifeSciences Corporation.
The Integra wave logo is a trademark of Integra LifeSciences Corporation.
©2006 Integra LifeSciences Corporation. All Rights Reserved.
Integra LifeSciences Services (France) SAS Immeuble Sequoia 2
97 allée Alexandre Borodine
Parc Technologique de la Porte des Alpes
69800 Saint Priest–France
Telephone: 33 (0) 4 37 47 59 10
Fax: 33 (0) 4 37 47 59 29