2 Nuvia HR-S Instruction Manual

Section 2

:

Technical Description

Table 1. Characteristics of Nuvia HR-S media.

Nuvia HR-S

Type of ion exchanger

Strong cation

Functional group

–SO3

Total ionic capacity

100–180 µeq/ml

Dynamic binding capacity*

>70 mg/ml at 300 cm/hr

Shipping counterion

Na+

Median particle size

50 ± 10 µm

Recommended linear flow

rate range

50–200 cm/hr

Chemical stability

1.0 N NaOH (20°C)

up to 5 weeks

0.1 N NaOH (20°C)**

up to 5 years

Compression factor

1.25 (settled bed
volume/packed bed volume)

pH stability**

2–14 short term
4–13 long term

Shipping solution

20% ethanol

Regeneration

1–2 M NaCl

Sanitization

0.5–1.0 N NaOH

Storage conditions

20% ethanol or 0.1 N NaOH

* 10% breakthrough capacity determined with 5.0 mg/ml human IgG in 20 mM

NaOAc pH 5.0
** Data derived under accelerated conditions at 60°C.