Nuvia HR-S Instruction Manual 9

Section 7

:

Regeneration and

Sanitization

After each run, the packed bed should be washed with 2–6

bed volumes of 1–2 M NaCl or until absorbance returns to

baseline to remove reversibly bound material. The column can

then be sanitized in 1.0 N NaOH at 50–100 cm/hr; a minimum

contact time of 120 min is recommended.

Section 8

:

Cleaning in Place

(CIP) and Sanitization

If a column no longer yields reproducible results, the

media may require thorough CIP and sanitization after

regeneration to remove strongly bound contaminants.

Acceptable CIP agents include 25% acetic acid, 8 M urea,

1% Triton X-100, 6 M potassium thiocyanate, 70%

ethanol, 30% isopropyl alcohol, 1 N NaOH, 6 M guanidine

hydrochloride, and 1 to 5% SDS.

Section 9

:

Storage

For long-term storage, Nuvia HR-S media should be
equilibrated with 0.01–0.1 N NaOH or 20% ethanol
(see Table 1).

Section 10

:

Regulatory Support

Regulatory support files are available for Nuvia HR-S

media. If you need assistance validating the use of Nuvia

media in a production process, contact your local Bio-Rad