beautypg.com

Bio-Rad Nuvia™ HR-S Media User Manual

Page 10

background image

6 Nuvia HR-S Instruction Manual

Section 5

:

Column Packing

Evaluation

When column packing is complete, equilibrate the

column until baseline conductivity is stable. To test the

effectiveness of column packing, inject a sample of a low

molecular weight, unretained compound (for example,

acetone or 1 M NaCl). If acetone is used as the test

marker (use a UV absorbance monitor set at 280 nm), the

equilibration buffer must have a salt concentration

<100 mM. If 1 M NaCl is the test marker (use a

conductivity monitor), then the equilibration buffer salt

concentration should be 100–200 mM. The sample

volume should be 1–3% of the total column volume.

Column testing should be operated using the same linear

velocity used to load and/or elute the sample.

To obtain comparable height equivalent to a theoretical

plate (HETP) values among columns, the same conditions

must be applied.

N = Number of theoretical plates
N = 5.54(V

e

/W

½h

)

2

L = Bed height, cm
HETP = L/N
V

e

= Peak elution volume or time

W

½h

= Peak width at peak half height in volume or time

V

e

and W

½h

should always be in the same units

Peaks should be symmetrical and the asymmetry factor as close
as possible to 1.

Peak asymmetry factor calculation:

A

s

= b/a

a = Front section of peak width at 10% of peak height
bisected by line denoting V

e

b = Latter section of peak width at 10% of peak height
bisected by line denoting V

e

A

s

equal to 0.8–2.5 should be easily achieved under normal

operating conditions.