Nuvia HR-S Instruction Manual 7

Section 6

:

Operation and

Maintenance

A linear flow rate of 100 cm/hr with a 20 cm bed is a

recommended starting point for Nuvia HR-S. Purification

may be optimized by changing the pH, flow rate, or ionic

strength of the elution buffer, modifying the gradient profile,

or experimenting with different buffer salts.

Note: Due to the smaller particle size of Nuvia HR-S

media, it is recommended to run the process at flow rates

no greater than 200 cm/hr to minimize backpressure and

headspace formation.

Figure 1 shows the effect of flow rate on backpressure.

Fig. 1. Nuvia HR-S media pressure/flow performance for a 20 cm

diameter x 20 cm bed height column packed to a compression
factor of 1.25.