beautypg.com

Bio-Rad Nuvia™ IMAC Resin User Manual

Page 8

background image

4 Bio-Scale Mini Nuvia IMAC Ni-Charged

Nuvia IMAC Cartridges are compatible with the aqueous buffers
most commonly used with IMAC purification techniques.

Table 3. Chemical compatibilities for Nuvia IMAC Ni-Charged

Cartridges.*

Reagent

Stability

Chelating agent

EDTA

1 mM

Sulfhydryl reagents

β-mercaptoethanol

20 mM

DTT

10 mM

TCEP

20 mM

Detergents

Nonionic

(Triton, Tween)

2%

Zwitterionic

(CHAPS, CHAPSO)

1%

Denaturing agents

Guanidine-HCl

6 M

Urea

8 M

Other additives

NaCl

2 M (include at least 300 mM NaCl in buffers)

MgCl

2

100 mM (use HEPES to prevent precipitation)

CaCl

2

5 mM (use HEPES to prevent precipitation)

Glycerol

20% (backpressure may increase significantly,

slower flow rates may be required)

Ethanol

20%

* Nuvia IMAC binding capacities are unaffected up to the concentrations given when employing
typical reagents used for histidine-tagged protein purification.