Bio-Rad Nuvia™ IMAC Resin User Manual
Page 10

6 Bio-Scale Mini Nuvia IMAC Ni-Charged
Section 4
Connection to Medium- and
High-Pressure Chromatography Systems
Bio-Scale Mini Cartridges can be connected to any medium- and
high-pressure liquid chromatography system set to a maximum
pressure limit of 45 psi (3 bar, 0.3 MPa). Bio-Rad offers two fitting
kits for easy connection of a Bio-Scale Mini Cartridge to medium-
or high-pressure chromatography systems.
NGC
™
Chromatography Systems and HPLC Systems
The Luer to 10-32 Adaptor Fittings Kit provides fittings necessary
to connect the cartridge to nut- and ferrule-type fittings found on
the NGC System and on most HPLC systems.
BioLogic DuoFlow
™
Systems
The Bio-Scale Mini Cartridge to BioLogic
™
System Fittings Kit
includes a 1/4-28 female to male luer and 1/4-28 female to female
luer to connect a Bio-Scale Mini Cartridge to Bio-Rad’s BioLogic
DuoFlow Chromatography System.
Fittings kit ordering information can be found in Section 12 of this
manual.