Bio-Rad Nuvia™ IMAC Resin User Manual
Page 6

2 Bio-Scale Mini Nuvia IMAC Ni-Charged
Section 2
Product Information
Bio-Scale Mini Cartridges are disposable, easy-to-use, prepacked
chromatographic cartridges supplied ready for use in convenient
1 and 5 ml sizes. Cartridges are available for a variety of
chromatographic techniques, including desalting ion exchange
(IEX), affinity chromatography (AC), mixed-mode, and hydrophobic
interaction chromatography (HIC). Refer to bio-rad.com for a
complete listing of items in the Bio-Scale Mini Cartridge product
line.
Bio-Scale Mini Cartridges are quickly connected to liquid
chromatography systems or luer syringes. The cartridges can be
used with any liquid chromatography system capable of setting a
high-pressure limit of 45 psi (3 bar, 0.3 MPa). Alternatively, built-in
luer fittings offer convenient connection directly to a syringe for
quick, one-step purification.
Nuvia IMAC Ni-Charged Resin is also available in bottles. Refer
to the ordering information in Section 12 of this manual. For more
information go to bio-rad.com.
Table 1. Bio-Scale Mini Nuvia IMAC Ni-Charged Cartridge
specifications.
Sizes
1 and 5 ml bed volumes
Dimensions
1 ml: 40 mm length x 5.6 mm inner diameter
5 ml: 40 mm length x 12.6 mm inner diameter
Maximum pressure
tolerance
45 psi
Recommended
flow rates
1 ml: 1–2 ml/min (240–480 cm/hr)
5 ml: 5–10 ml/min (240–480 cm/hr)
Maximum flow rate
1 ml: 6 ml/min (1,440 cm/hr)
5 ml: 20 ml/min (963 cm/hr)
Fittings
Female luer inlet and male luer outlet
Column material
Polypropylene
Frit material
Polyethylene (HDPE)
Shipping conditions
20% ethanol
Storage
recommendations
20% ethanol
Autoclavability
Not autoclavable