Fig. 3. Flush phase. (Check valve is open.)

Section 2
Installation

2.1 Unpacking

The DynaLoop is shipped fully assembled as shown in Figure 1.

List of components

The shipping carton contains:

DynaLoop sample loop

Fittings and tubing to make all necessary plumbing
connections

DynaLoop spare parts

Instruction manual

2.2 Precautions and Recommendations

This section contains important information concerning the safe operation of this

accessory. Please take a few minutes to review it and observe all of the cautions and
warnings. Failure to do so can result in damage to the DynaLoop and/or the BioLogic
DuoFlow system, and void the warranty.

Maximum Operating Pressure

The maximum operating pressure of the DynaLoop is 1,000 psi (7 MPa, 70 atm). Set
the pump high pressure limit to a pressure below 1,000 psi.

The DynaLoop has an external plastic sleeve which protects the internal parts from
accidental damage. Should the glass tube break, the shield will contain all glass fragments.

Warning: Know the physical properties of the solvents you use. Refer to the Material
Safety Data Sheet that accompanies each solvent.

O-ring seals

Check valve

(open)

Sample end
connector

Sliding seal assembly

in contact with sample

end connector

BUFFER

SAMPLE