Table 2. dynaloop replacement parts – Bio-Rad BioLogic Sample Loading Loops User Manual
Page 19
Section 5
Appendix A Replacement Parts
Table 2. DynaLoop Replacement Parts
Product
Description
Catalog Number
DynaLoop Parts Kit, includes end cap O-rings, sliding
750-0450
seal O-ring, filter, collet/nut fittings, 1/8” ferrules,
1/4-28 nuts and ferrules, and 1/8”
PTFE tubing.
Luer to BioLogic System Fittings Kit,
732-0113
FEP Tubing, 1/8” pre-pump connections, 15’
750-0603
Section 6
Appendix B Solvent Compatibility
The O-rings and check valve ball, which are made from nitrile rubber (Buna-N), are
compatible with alcohols, aqueous buffers, and dilute aqueous acids and bases.
Typical HPLC solvents, such as dioxane, acetone, organic esters, ethers,
dimethylformamide, tetrahydrofuran, and concentrated acids and bases, can swell the seal
and check valve ball. Test for compatibility with questionable solvents by immersing these
components in a beaker of solvent overnight. Then examine for stability.
Luer-Lok is a registered trademark of Becton, Dickson, and Company.
Milli-Q is a registered trademark of Millipore Corporation.
PEEK is a registered trademark of Victrex PLC.
Ryton is a registered trademark of Chevron Phillips Chemical Company LLC.
Tefzel
is a registered trademark of the E.I. Du Pont de Nemours Co.
15