Bio-Rad Bio-Plex Pro Human Acute Phase Reagent and Diluent Kits User Manual

Page 31

background image

Acquire Data
1. Shake the assay plate at 900 rpm for 30 sec immediately before

acquiring data. Failure to do so will result in increased data

acquisition time due to bead settling.

2. Visually inspect the plate and ensure that the assay wells are filled with

buffer prior to placing the plate in the Bio-Plex microplate platform.

3. Slowly remove the sealing tape and any plate cover before placing

the plate in the reader.

4. Select Step 7 (Run Protocol):

a)

Choose

100 beads per region.

b)

In Advanced Settings, set the bead map to

25 region.

c)

In Advanced Settings, set the sample size to

50 µl.

d)

In Advanced Settings, confirm that the DD gate are set to

5,000

(low) and

32,000 (high).

NOTE: These settings are already selected in preset

downloaded protocols.

e)

Select Start and save the protocol (.pbx) file. Then follow the

instructions for data acquisition.

5. If acquiring data from ore than one plate, empty the waste bottle and

refill the sheath bottle after each plate (if HTF not present). Select Wash

Between Plates

and follow instructions for fluidics maintenance.

Then repeat the Prepare Protocol and Acquire Data steps.

NOTE: Use the Wash Between Plates command after every plate run

to reduce the possibility of clogging the instrument.

6. When data acquisition is complete, select Shut Down

and follow

the instructions.

27